48T
$299.00
96T
$399.00
Species:
Human
Alternative Name:
Androstenolone sulfate; Prasterone sulfate; Androst-5-en-3β-ol-17-one 3β-sulfate
Assay Type:
Competitive
unit:
ng/mL
Standard:
1000ng/mL
Sensitivity:
9.38 ng/mL
Detection Range:
15.63-1000ng/mL
Assay Time:
1.5H
Sample Type:
Serum, Plasma, Tissue homogenate and Other biological samples;Sample Volume=50μL
Research:
Enzyme & Kinase;
Test Principle:
This ELISA kit uses the Competitive-ELISA principle. The micro ELISA plate provided in this kit has been pre-coated with Human DHEA-S. During the reaction, Human DHEA-S in the sample or standard competes with a fixed amount of Human DHEA-S on the solid phase supporter for sites on the Biotinylated Detection Ab specific to Human DHEA-S. Excess conjugate and unbound sample or standard are washed away, and Avidin-Horseradish Peroxidase (HRP) conjugate are added to each micro plate well and incubated. Then a TMB substrate solution is added to each well. The enzyme-substrate reaction is terminated by the addition of stop solution and the color turns from blue to yellow. The optical density (OD) is measured spectrophotometrically at a wavelength of 450 nm ± 2 nm. The concentration of Human DHEA-S in tested samples can be calculated by comparing the OD of the samples to the standard curve.
| ng/mL | OD | Corrected |
|---|---|---|
| 1000 | 0.296 | |
| 500 | 0.396 | |
| 250 | 0.59 | |
| 125 | 0.851 | |
| 62.5 | 1.146 | |
| 31.25 | 1.541 | |
| 15.63 | 1.855 | |
| 0 | 2.519 |
Intra-assay Precision (Precision within an assay): 3 samples with low, mid range and high level were tested 20 times on one plate, respectively.
Inter-assay Precision (Precision between assays): 3 samples with low, mid range and high level were tested on 3 different plates, 20 replicates in each plate, respectively.
| Intra-assay Precision | Inter-assay Precision | |||||
|---|---|---|---|---|---|---|
| Sample | 1 | 2 | 3 | 1 | 2 | 3 |
| n | 20 | 20 | 20 | 20 | 20 | 20 |
| Mean(ng/mL) | 18.15 | 212.5 | 473.21 | 50.96 | 218.66 | 288.83 |
| Standard deviation | 0.55 | 9.1 | 21.08 | 3.05 | 10.46 | 21.88 |
| C V (%) | 7.03 | 7.04 | 4.78 | 5.94 | 6.27 | 2.89 |
| Sample Type | Range(%) | Average Recovery(%) |
|---|---|---|
| Serum (n=8) | 86-99 | 92 |
| EDTA plasma (n=8) | 90-105 | 97 |
| Cell culture media (n=8) | 80-95 | 87 |
| Serum (n=5) | EDTA plasma (n=5) | Cell culture media (n=5) | ||
|---|---|---|---|---|
| 1:2 | Range (%) | 93-105 | 92-105 | 87-96 |
| Average (%) | 99 | 102 | 94 | |
| 1:4 | Range (%) | 85-97 | 92-105 | 96-101 |
| Average (%) | 97 | 103 | 97 | |
| 1:8 | Range (%) | 90-99 | 85-101 | 85-92 |
| Average (%) | 92 | 89 | 90 | |
| 1:16 | Range (%) | 82-101 | 85-94 | 80-96 |
| Average (%) | 98 | 86 | 92 |